Development and implementation of a wireless real-time radiation monitoring system for LINAC beam delivery monitoring
Vo Hong Hai, Phuc Tri Toan Nguyen and Hieu Trung Nguyen
Science & Technology Development Journal 2024, 27():1-7
Abstract:
Introduction: In this work, we developed a wireless, remote, real-time radiation monitoring system designed to oversee beam delivery in a radiation therapy room equipped with a medical linear accelerator (LINAC).
Methods: This system utilizes a Geiger-Müller detector paired with embedded electronic hardware to accurately record radiation count rates in real-time. The data collected by this system are transmitted through LAN/WAN networks to the internet, ensuring instantaneous accessibility. In addition, a web server and mobile application were developed to display, receive, and archive the data from the radiation counter.
Results: Our system was deployed in the radiation therapy room of the Oncology Hospital in Ho Chi Minh City, demonstrating a remarkable data reception rate of up to 99.8% over a three-day test period from June 27th to 29th, 2022. The system effectively identified beam-on instances and provided precise measurements of the number and duration of beam-on events.
Conclusion: This study demonstrates the feasibility of remote real-time radiation monitoring in medical settings and highlights the potential for enhancing radiation safety and treatment efficacy in external beam radiotherapy.
More detail >>